SpiroStar DX
World's smallest Windows
compatible spirometer
(Product code M921)
 |
New universal SpiroStar spirometer represents
the state of the art in dynamic spirometry. It is compact weighting only 22 g. SpiroStar
can be connected into a standard (RS-232C) serial port of a computer. It will save all the
valuable working space of your desk for you. SpiroStar is accurate and precise and it
complies with the ATS spirometry standards (1994) giving you the assurance of reliable
results. Heuristic system algorithm monitors the stability of the system and guarantees
repeatable and reliable session results.
|
The application software is Windows (95 / 98
/ NT 4.0/ 2000 / XP) compatible Spiro2000, which provides easy operation, flexible
parameter selection and comprehensive choice of report formats.
You can modify the reports
to suite your requirements. Session results can be transferred into hospital information
systems via local area network or emailed to consulting physician fast and faultlessly.
|
Calibration database provides a tool for
professional quality assurance and management system.
SpiroStar includes an automatic
temperature sensor to monitor and correct variations in operating environment.
SpiroStar
uses disposable Medikro Flow Transducer, which provides for better patient safety by
eliminating potential contamination.
It increases the system throughput and saves
operating costs.
The instrument and the system are designed to assist you in developing
and maintaining a better and safer lung function testing system for your patients.
|
Technical
specificationsSystem components:
- Personal computer (minimum requirements):
- Pentium 433 Hz, Windows 95
/ 98 / NT 4.0 / 2000 / XP, 32MB/32MB/64MB/128MB, 800x600 display with 256 colours, CD-drive
keyboard and
mouse
- IE5 and 32-bit ODBC drivers (included in Spiro2000
installation kit)
Personal computer (recommended):
- Pentium 1GHz, Windows 95 /
98 / NT 4.0 / 2000 / XP, 64MB/64MB/128MB/256MB/512MB, 800x600 display with
true colours, CD-drive, keyboard and
mouse
- IE5 and 32-bit ODBC drivers (included in Spiro2000
installation kit)
Graphics printerSpiroStar DX Spirometer (including):
- SpiroStar DX Spirometer unit (M9453)
- Disposable Flow Transducer (M9228)
- Tubing (M9220)
- Spiro2000 Spirometer Software (M9831)
- Quick Instructions Guide and Installation Guide
- Users Manual (electronic)
SpiroStar
DX Spirometer Unit
- Physical dimensions: 55 mm x 34 mm x 16 mm
- Weight: 22 g
- Frequency response: 0
30 Hz (-3 dB)
- Supply voltage: via signals in RS-232-C serial port
- Temperature correction: software based
- Temperature range (operating): + 10 ° C
+ 40 ° C
- Temperature range (storage): - 10 ° C
+ 50 ° C
- Relative humidity range: 5 %
85 %
- Pressure range: 600
1060 hPa
- Safety standards : EN60601-1, Class 2a medical device (MDD,
93/42/EEC), CE certified
Disposable Flow Transducer – the best solution to patient safety
- Screen-type interchangeable disposable pneumotach
- Material polyethylene, can be disposed with incineration
without toxic emissions
- Flow range - 14 l/s
+ 14 l/s
- Back pressure: less than 0,6 kPa/(l/s), meets ATS, ACCP &
ECCS standards
- Accuracy: +- 3% without calibration, +- 2% with volume
calibration
- Sterilisation not required, the entire Flow Transducer is
disposable
- Outlet port adapter to fit oval end of the flow transducer
with round syringe outlet port included
Spiro2000 - Spirometer software for the professionals Operating System:
Windows 95 / 98 / NT 4.0 / 2000 /
XP Calibration
- Calibration with standard 3000 ml pulmonary calibration
syringe, user defined volumes allowed
- Calibration results are recorded and stored in the calibration
log file
Measurement
- Collection of whole signal data, collection rate: 100 Hz
- Max duration of one measurement: 60 s
- Measurement protocols: TV, VC, FEVC, FIVC,
FVC and MVV
- Meets ATS/ERS standards
- Determination of 0-time in FVC by back extrapolation
- Metric and US units
- Calculated variables:
- FEVC, FEV0.5, FEV1, FEV6, FEV0.5%(VC),
FEV0.5%(VCFVC), FEV0.5%(FEV6), FEV0.5%(FEVC), FEV1%(VC), FEV1%(FEVC),
FEV1%(VCFVC), FEV1%(FEV6), PEF, PEFLPM, MEF75, MEF50, MEF25, MMEF(FEF25-75), METT, AEFV,
FET and VEXT
- FIV6, FIV0.5%(VC), FIV0.5%(VCFVC), FIV0.5%(FEV6), FIV0.5%(FIVC), FIV1%(VC), FIV1%(FIVC), FIVC,
FIV1%(VCFVC), FIV1%(FIV6), PIF, PIFLPM, MIF75, MIF50, MIF25, MMIF(FIF25-75), MITT, AIFV,
FIT and VIEXT
- VC
- TV, FR, MV
- MVV, MVVFR and MVVT
Result analysis
- Criteria of best t results:
- Maximum of FVC+FEV1 (or FVC+FIV1), ATS guidelineEnvelope of F/V curves, ERS guideline
- Maximum area under F/V curve
- Automatic interpretation of bronchodilatation test
- Several interactive graphs and tables. Detailed viewing by
zoom, signal change by check boxes. Trending.
- Reference values: several pre-defined national sets. Reference
values also in a histogram form.
- Final report: Page size A4/Letter, HTML based, user defined,
several pre-defined formats available.
- Database analysis: by MS-Access or any other ODBC- and
access-capable database programs. Data transfer to Excel (MSExcel not
included)
Remote
use
- Database access via local area network (LAN)
- Session data transfer via Email
|